• BS-800 Auto Clinical Chemistry Analyzer

    Discrete, random access, fully automated 800 tests per hour, up to 1200 tests per hour with ISE Up to 68 on-board chemistries & 3 ions Onboard capacity of 140 sample positions 165 permanent used glass cuvettes 100µl minimum reaction volume Indirect method ISE module with higher throughput Automatic probe cleaning, liquid level detection, clot & bubble detection and collision protection Water circulation cooling for reagent and sample carousels 8-step auto washing system with pre-warmed water 3-head mixers for sample and reagent respectively Automatic sample dilution for abnormal sample Reagent & sample continuous loading and offloading Water quality monitor Bidirectional connection to LIS host Optional Remote Maintenance System (RMS) Integrated with original chemistry reagents, calibrators and controls

Throughput:
800 photometric tests per hour for a standalone unit 
and single analytical unit of a modular system, up to 
1200 tests per hour with ISE 
From 800 to 2400 tests per hour for a modular system 
with dierent congurations
On-board tests: 
68 photometric tests + 3 ISEs + 3 serum indices
Sample Handling
Sample tray: 
140 positions, including 25 cooled 
positions for calibrators and controls
SDM:
300 samples by 30 racks
Sample volume:
1.5~35 µL, step at 0.1µL
Sample probe:
Liquid level detection, clot detection, and collision protection
Reagent Handling
Reagent tray: 
120 positions in coaxial disk for R1, R2, R3 and R4
Reagent volume: 
15~300 µL, step at 0.5 µL
Reagent probe:
Liquid level detection, bubble detection, and collision protection
Built-in Bar Code Reader (optional)
Sample and reagent bar code reader’s support 
Codabar, ITF (Interleaved Two of Five), Code128, 
Code39, UPC/EAN, and Code93; 
Capable to connect with LIS in the bi-directional mode
Reaction System
Reaction volume: 
100~360 µL
Reaction temperature: 
37˚C with 0.1˚C fluctuation
Reaction cuvettes:
165 glass cuvettes with 8-step auto wash
Optical System
Light source: 
Tungsten-halogen lamp
Photometer: 
Grating photometer
Wavelength:
340nm, 380nm, 412nm, 450nm, 505nm, 546nm, 
570nm, 605nm, 660nm, 700nm, 740nm, 800nm
Absorbance range: 
0~3.4Abs (10mm conversion)
ISE Module (Optional)
Principle:
Indirect K+, Na+ and Cl- , with 22 µL sample aspiration
Control and Calibration
Calibration mode:
Linear (one-point, two-point, and multi-point), 
Logit-Log 4P, Logit-Log 5P, Spline, 
Exponential, Polynomial, Parabola
Control rules:
Westgard multi-rule, Twin plot
Operation Unit
Operation system: 
Windows® XP Professional/Home SP2 or above, 
Windows® 7, Windows® 8, Windows® 10
Scalability
Upgradable to SAL 6000 modular system, a seamless 
integration of clinical chemistry module and 
chemiluminescence immunoassay module