C3510 Auto Coagulation Analyseur
Mechanical method (clotting); Optical method (chromogenic, immunoturbidimetry) Test items (For specific reagents) Clotting items: PT, APTT, Fib, TT, ProC, ProS, coagulation factors ( Ⅷ, Ⅸ), etc. Chromogenic items (405nm): ATⅢetc. Immunoturbidimetry items (575nm): D-dimer, FDP, etc.
Principles
Mechanical method (clotting); Optical method (chromogenic, immunoturbidimetry)
Test items (For specific reagents)
Clotting items: PT, APTT, Fib, TT, ProC, ProS, coagulation factors ( Ⅷ, Ⅸ), etc.
Chromogenic items (405nm): ATⅢetc.
Immunoturbidimetry items (575nm): D-dimer, FDP, etc.
Throughput
PT: Up to 300 test/h
D-dimer: 100 test/h
Testing channels
Clotting channels: 4
575nm optic channel: 5
405nm optic channel: 1
Barcode
Built-in barcode scanner
Sample position
80 positions, random access
STAT position
3
Reagent position
24 (cooling) + 4 reagent position
5 positions with stirring function
Automatic cuvette loading
1000 cuvettes autoloading
Sampling system
individual reagent and sample probe.
Both probes are equipped with liquid-level detection functions and collision protection in the vertical direction.
Reagent probe equipped with heating function.
Operating environment
Temperature: 10℃~30℃; Relative humidity: ≤70%
Power supply
AC100-240V, 50/60 Hz