  • BS-230 Auto Clinical Chemistry Analyzer

    Discrete, random access, fully automated, bench-top Up to 200 tests per hour, up to 400 tests per hour with ISE Optical system: 8 wavelengths 100µl minimum reaction volume Independent mixing bar Up to 80 positions for reagent, Up to 80 positions for sample 2~12℃ cooling compartment for reagent and sample 40 disposable reaction cuvettes Sample volume: 2μL~45μL, step by 0.1μL, reagent volume: 10μ L~250μL, step by 0.5μL Automatic probe cleaning for interior and exterior Liquid level detection and collision protection for probes Bi-directional connection to LIS host Original system pack reagent ready to use

PDF Татах

System function:
Automatic, Discrete, Random Access, Bench-top
STAT sample priority
Throughput
Up to 200 tests/hour, up to 400 tests/hour with ISE
Measuring principles: 
Absorbance photometry, Turbidimetry, Ion
Selective Electrode technology
Methodology: 
End-point, Fixed-time, Kinetic, optional ISE,
Single/Dual/ reagent chemistries, 
monochromatic / bi-chromatic
Original system pack reagent ready to use
Close system and open system is optional
Reagent/Sample Handling
Reagent/Sample tray:
80 positions for reagents and 40 positions 
for samples in the 24-hour refrigerated 
compartment (2~12℃)
Reagent volume: 
10~250μl, step by 0.5μl
Sample volume: 
2~45μl, step by 0.1μl
Reagent/Sample probe: Liquid level detection, vertical collision 
protection and inventory checking, reagent pre-warming
Probe cleaning: 
Automatic washing for interior and exterior 
Carry over < 0.05%
Probe cleaning: Automatic washing for interior and exterior Carry over < 0.05%

Internal bar code reader (optional)
Capable to communicate with LIS in bi-directional mode

Reaction System:
Cuvette: 40 disposable cuvettes 
Reaction volume: 100~360μl 
Operating temperature: 37℃±0.1℃
ISE Module (optional) 
Measuring K+, Na+, Cl-
Mixing Unit:
Independent mixing bar
Optical System
Light Source:
Halogen-tungsten lamp
Wavelength:
8 wavelengths, 340nm、405nm、450nm、 
510nm、546nm、578nm、630nm、670nm
Absorption range: 
0~4.0 Abs (10mm conversion), resolution 0.0001Abs
Stray Light: 
5.6Abs
Control and Calibration
Calibration modes: 
Linear (one point, two points and 
multi-points), Logit-Log 4P, Logit-Log 5P, 
spline, exponential, polynomial, parabola
Control Rules:
X-R, L-J, Westgard multi-rule, Cumulative sum 
check, twin plot
Operation Unit
Operation system: 
Windows 8
Interface: 
RS-232
Working Conditions
Power Supply:
200~240V, 50/60Hz, ≤1000VA or 100~130V, 60Hz, ≤1000VA
Dimension:
690 mm (length) ×580 mm (depth) ×595 mm (height)
Weight:
47 kg
Water Consumption:
≤ 2 L/H